Rhythm
Strip
Recorder
Procedure
To replace the batteries in the Rhythm Strip Recorder, follow the
procedure for the Blood Pressure Monitor on page 3-9, except for Step 7.
The Recorder will make a series of beeps.
After Step 6:
Step 7. Measure your heart rhythm with the Rhythm Strip Recorder to
verify that it is working properly.
Note
If the Rhythm Strip Recorder does not seem to be operating properly or
the Radio Test fails, contact your healthcare provider.
Replace the batteries in the Rhythm Strip Recorder every 6 months
or sooner if required.
If the Rhythm Strip Recorder beeps twice quickly before beginning
normal on and off beeping, the batteries are low and should be
replaced within a week.
The Rhythm Strip Recorder uses 4 AA alkaline batteries.
If the series ends with a rising tone, the batteries have been inserted
correctly.
If the series ends with a falling tone or no tone, the batteries have
been inserted the wrong way. Repeat the battery insertion steps.
Replacing Batteries
3-11
Maintenance