Electromagnetic Compatibility (M3813B Scale)
Immunity
IEC 60601Test
Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
5-16
Specifications and Regulatory
Electromagnetic Immunity
Compliance
Level
Level
Not Applicable
3 V/m
Electromagnetic Environment
Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the M3813B Scale, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d
=
1
2 .
P
d
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter's specified
output power and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol :
=
2
3 .
P
b