Descargar Imprimir esta página

Emissions And Immunity; Reducing Electro-Magnetic Interference; Electromagnetic Emissions - Philips Patient Telemonitoring Set Instrucciones De Uso

Publicidad

Idiomas disponibles

Idiomas disponibles

Electromagnetic Compatibility (M3813B Scale)
Emissions
The M3813B Scale may be susceptible to interference from other RF
and
energy sources. Examples of other sources of RF interference are other
Immunity
medical electrical devices, cellular products, information technology
equipment, and radio/television transmission. If interference is
encountered, as demonstrated by anomalous product behavior, attempt to
locate the source. Assess the following:
Reducing
Once the source is located, attempt to attenuate the interference by
Electro-
distancing the product from the source as much as possible. If assistance
magnetic
is needed, contact your local Service Provider.
Interference
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Note
The M3813B Scale is intended for use in the electromagnetic
environment specified above. The customer or the user of the M3813B
Scale should assure that it is used in such an environment.
5-14
Specifications and Regulatory
Is the interference intermittent or constant?
Does the interference occur only in certain locations?
Does the interference occur only when in close proximity to certain
equipment?

Electromagnetic Emissions

Compliance
Electromagnetic Environment Guidance
Group 1
The M3813B Scale uses RF energy only for its internal
function. Therefore, its RF emissions are very low and not
likely to cause interference in nearby electronic equipment.
Class B
The M3813B Scale is suitable for use in all establishments,
including domestic and those directly connected to the
public low-voltage power supply network that supplies
Not Applicable
buildings used for domestic purposes.
Not Applicable

Publicidad

Capítulos