The VERSAJET II Hydrosurgery System (66800039) is intended for use in the electromagnetic environment specified below. The customer or the
user of the VERSAJET II Hydrosurgery System should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
WARNING: The VERSAJET II Hydrosurgery System should not be used adjacent to or stacked with other electrical equipment and that if adjacent
or stacked use is necessary, the VERSAJET II Hydrosurgery System should be observed to verify normal operation in the configuration in which it
will be used.
Recommended separation distances between portable and mobile RF communications equipment and The VERSAJET II Hydrosurgery
System (66800039).
The VERSAJET II Hydrosurgery System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of The VERSAJET II Hydrosurgery System can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and The VERSAJET II Hydrosurgery System as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Guidance and Manufacturer's declaration – electromagnetic emissions.
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
150 kHz to 80 MHz
d = 3.5√P
d = 12√P
0.35
1.2
1.1
3.8
3.5
12
11
38
35
12
Compliance
Electromagnetic environment - guidelines
The VERSAJET II Hydrosurgery System uses RF energy only
Group 1
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause interference in nearby
Class B
electronic equipment.
Class B
The VERSAJET II Hydrosurgery System is suitable for use
in hospital establishments and those directly connected to
the public low-voltage power supply network that supplies
Complies
buildings used for industrial purposes.
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
19
19
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23
EN