Power display and fault indicators
1.
Power display – displays power setting
Footswitch
5
4
1
2
01 = lowest power setting
10= highest power setting
3
2
1
3
Fault indicators – illuminate if a fault is present
2.
Footswitch not connected indicator
– an amber LED will illuminate when the footswitch
is not fully connected or is a defective footswitch
3.
System fault indicator
– a red LED will illuminate indicating a power fault
or an out of tolerance pressure condition
• If a fault occurs, the console should be sent
back to the manufacturer for service. DO NOT
attempt to open the unit to perform repair.
1.
Footswitch pedal – depressing the footswitch
pedal activates the cutting jet
2.
UP toe button – depressing the toe button next to
the UP arrow increases the power setting
3.
DOWN toe button – depressing the toe button
next to the DOWN arrow decreases the power
setting
4.
Footswitch connector – connects the footswitch
to the footswitch socket on the console
5.
Footswitch cable – connects the footswitch
connector to the footswitch
7
7
7
EN
EN
EN