Physical appearance check - Caution: console is unplugged and placed on bench for this step
Step
1
Check console for visible dents, cracks and missing pieces.
2
Check that ventilation slots are free from obstructions, dust or foreign materials.
3
Check that labels and markings are intact and legible with no smudges.
Check that the pump interface rotates between lock and unlock positions and
4
remains in place.
5
Check footswitch and attached cable for signs of damage or missing insulation.
6
Check footswitch mechanicals for system activation by depressing and releasing.
7
Inspece UP/DOWN power controls for excessive wear.
Inspect power cord for any frayed or missing insulation and bent, loose or
8
missing plug blades or earth/ground.
System diagnostics check - Note: console is plugged in and placed on bench for this step
9
Adjust main power switch to ON/OFF positions; verify illumination status when ON.
10
Verify that system indicator(s) are operating properly.
10a
Verify that the footswitch connection LED indicator flashes amber ON/OFF
10b
Verify that the system fault LED indicator flashes red ON/OFF
10c
Verify that the pump interface illuminated green light ring flashes green ON/OFF
10d
Verify that the power display flashes 88 ON/OFF
11
Verify console defaults to power setting 01 when unit is turned ON.
12
Verify system fault indicator red LED does not illuminate.
Verify the console power display value changes when arrows are depressed –
13
there are 10 settings.
14
Verify that the pump interface illuminated green light ring functions properly.
15
Verify that the footswitch connection LED indicator functions properly.
Console speed check - Note: console is plugged in and placed on bench for this step
• A hand held optical tachometer is needed for this step
16
Power ON console by pressing illuminated power switch.
17
Insert the speed stick into the pump interface and turn to the LOCKED position.
18
Set console to its lowest power setting 01.
19
Depress footswitch pedal allowing motor and transmission to cycle.
20
Measure speed per console speed check instructions; verify the RPM is 425 ± 50.
21
Set console to its highest power setting 10.
22
Depress footswitch pedal allowing motor and transmission to cycle.
23
Measure speed per console speed check instructions; verify the RPM is 1290 ± 100.
Electrical equipment safety check - Note: console is plugged in and placed on bench for this step
• A meter capable of measuring micro amps is required
24
Power ON console by pressing illuminated power switch.
25
Attach test leads to ground plug and pump interface as per electrical testing instructions.
Take readings as per electrical testing instructions to check if leakage is occurring
26
(Table 2).
27
Disconnect test leads and power OFF by pressing illuminated power switch.
28
Unplug power cord from electrical outlet and from rear panel of console.
Summary of results
29
Overall physical appearance showed no anomalies and unit is clean.
30
Basic functional setup indicates functions evaluated are operational.
31
Console speed verification meets specification as defined in test protocol.
32
Electrical equipment safety testing indicates readings meet requirements.
• Please photocopy this page when performing the checklist so that you will always have a blank copy. • If any of the Summary
Results are "NO" this could be an indication of a need for Service or Repair. • Please contact your Smith & Nephew Customer Care
representative immediately at 1-800-876-1261 (USA only). • Have this report available at time of call.
Yes
No