Figure 16 – The touch screen – Anesthetic Tab
11.
Following the procedure, the
deploy a marker through the Probe. Enter Marker Mode by pressing the "Marker" tab on the touch
screen as illustrated in Figure 17. Deploy the marker per the marker Instructions for Use.
Rotate the sample notch 180° by pressing the "180°" button on the touch screen as illustrated in
Figure 17. Exit Marker Mode by pressing the "Sample" button on either the handswitch or
Foot Pedal.
Figure 17 – The touch screen – Marker Tab
After Use
Note: The system may be left on if additional procedures are to be performed; it is not
recommended to leave the main power on overnight. It is not recommended to unplug the
Driver after each use. The Driver and Foot Pedal may be left connected to the
Breast Biopsy System if all procedures are to be performed with the same Driver.
®
13.
To set the system to standby, press and hold the switch on the front of the console (illustrated in
Figure 12
located on the rear of the console (illustrated in Figure 11

Cleaning and Maintenance

To clean the E
E
®
E
®
Breast Biopsy System, cables, and Foot Pedal with a soft, lint-free cloth dampened with
Dispatch or a similar cleaning solution or disinfectant. Follow the procedures approved by your facility or
of the cleaner or disinfectant used.
Biopsy System or cables. Spraying the system may cause it to malfunction and will void the warranty.
Do not submerge any component of the E
may damage the electrical components of the device. If the system is inappropriately cleaned, it may
cause the system to malfunction and will void the warranty. Do not autoclave any components of the
E
E
®
Disposal of Equipment
disposal of the E
E
Breast Biopsy System or any of the reusable accessories at the end
®
of the E
E
Breast Biopsy System's useable life per Waste of Electrical and Electronic
®
Instructions for Use for the disposable device.
Do not dispose of the E
E
®
Breast Biopsy System by placing into trash receptacles.
®
Breast Biopsy System allows the operator to
180° button
E
E
®
Routine Maintenance
No calibrations or adjustments are necessary for the E
normal use. In addition, the following routine maintenance activities may be completed:
• Electrical safety testing should be performed at intervals no greater than 12 months, using a
standard medical safety analyzer.
• Regularly inspect all cables and connectors for possible wear or damage.
Odor Control Filter Replacement
Fuse Replacement
rear panel of the console. Using a small slotted screwdriver, press the release tab located near the
center of the Fuse Holder. Pull the Fuse Holder assembly outward to remove it. Install the Fuse Holder.
Remove and replace any blown fuses in the Fuse Holder assembly and snap it back into place. The
Fuse Holder orientation determines the voltage. Rotate the fuse 180 degrees for 230V orientation.
Recommended Fuse for Replacement:
Troubleshooting
The E
E
®
Breast Biopsy System will display any alert conditions that occur. If the system
can identify the cause of the alert, the information is displayed on the touch screen.
conditions, when corrected, allow the system to recover automatically and return to a ready state. Other
alerts require pressing a button on the touch screen to clear. If a Driver is connected and is the cause of
the alert, follow the Driver Instructions for Use to clear the alert.
Power Switch on the rear of the system (illustrated in Figure 11
back on, and restart the E
the console (illustrated in Figure 12
If any alert condition persists after correcting known problems and clearing the alert condition from the
screen, contact Bard for service.
®
Breast
3
E
®
Breast Biopsy System outside of
Figure 18.
Figure 18 – Odor Control Filter
E
®
Breast Biopsy System with the standby button on the front of
E
Breast
®