multipiezo
8.5
Cleaning Check
8.5.1
Required Materials
• Light source;
Once the cleaning operations have been
completed, check the scaler handpiece
and scaler nose cone under an adequate
source of light, if necessary using a
magnifying glass 2.5X, paying attention
to any parts that may conceal soil residue
(threading, cavities, grooves) and, if need
be, repeat the cleaning cycle if soil is still
visible. Finally, check the integrity of those
parts and those elements that could have
deteriorated during use;
Repeat the control operations for the
other accessories (inserts, tightening
wrenches), repeating if necessary the
cleaning cycle.
52
• 2.5X Magnifier.
1
2