Remove the two tubes of the pump from
the respective clutches positioned below
it;
Extract the peristaltic pump from its base,
pulling it towards yourself;
Attach the new peristaltic pump onto the
base until you hear it click in, and connect
the two pump tubes to the respective
couplings underneath it;
Refit the plastic protection onto the
peristaltic pump.
2
3
4