Remove the handpiece, nose cone, and
light pipe from the enzymatic solution
and gently brush their surfaces using a
brush with soft nylon bristles under hot
(40 °C ±5 °C) running water, for at least
10 minutes;
Place the scaler handpiece, nose cone,
and light pipe in the ultrasonic tank
immersed in demineralised water at
40 °C ±5 °C and run a cycle for at least
10 minutes;
Rinse the inner channel of the scaler
handpiece using
a 20
previously filled with demineralised
water. Repeat five times;
Dry the surface of the scaler handpiece,
nose cone, and light pipe with a clean,
non-abrasive, low-lint cloth.
CLEANING AND STERILISATION
8
9
10
ml
syringe
11
x5
H₂O
H₂O
45