multipiezo
CAUTION: Wait for the scaler handpiece
to cool down completely before use.
8.7.2.1
Special Information
Sterilisation parameters, in steam autoclave, used in Great Britain:
• Temperature: 134° C;
9
MAINTENANCE
If the appliance is not used for a long time
observe the following recommendations:
1.
Run a complete cleaning cycle on the
irrigation circuit using the "pump/flush"
function (see
Chapter 6 on page 27
2.
Empty the circuits of any residual water,
removing the tank and making the
handpiece work for a few seconds;
3.
Disconnect the appliance from the
mains;
4.
In case of long periods of non-use,
place the appliance back in its original
9.1
Replacing the Peristaltic Pump
On the back of the appliance there is the
plastic cover which covers the housing
of the peristaltic pump. Remove this
protection by pressing on the sides and
pulling towards yourself;
CAUTION: Before carrying out
operations on the peristaltic pump,
make sure the appliance is unplugged
and that the liquids container is not
connected.
56
• Time: 3 minutes.
packaging, in a safe place;
5.
Before using the appliance again, clean
and sterilise the handpiece, inserts, and
wrench following the instructions in
);
Chapter 8 on page 31
6.
Check that the inserts are not worn,
deformed or broken, with particular
attention to the integrity of the tip.
WARNING: Periodically check the
integrity of the power supply cable; if
damaged, replace it with an original Mectron
spare part.
1
;