Disconnect the irrigation solution tank
from the device;
Unscrew the cap from the irrigation
solution tank and drain the remaining
liquid;
Rinse the irrigation solution tank with
running water;
With the tank disconnected, run a flush
cycle (see points 7 to 10) to empty the
irrigation circuit of the liquid inside;
Fill the irrigation solution tank with 300ml
of distilled water;
NOTE: Use of distilled water is
recommended for long periods of
non-use. If the device is used after the
disinfection procedure, use common
running water.
Holding the irrigation solution tank in a
vertical position, insert it on the machine
core unit of the device until it comes into
contact;
CAUTION: Do not overturn the tank,
the cap is not watertight. The outflow
of physiological solution or aggressive
liquids may cause damage to the
surfaces.