so that the profile corresponds to the
one printed on the card. The printed
profile on the card has a red line
indicating the limit of wear;
• If the insert is shorter than the limit of
wear, its performance will be significantly
inferior compared to that of a new insert,
and should therefore be replaced.
If the layer of titanium nitride (gold-plated
surface), where present, is visibly worn, the
insert must be replaced. The use of a worn
insert reduces its efficiency.
Diamond Inserts: diamond inserts must be
replaced when the layer of titanium nitride is
visibly worn and in any case after a maximum
of 10 treatments.
When the nitriding is consumed, the
cutting edge loses effectiveness; a possible
regrinding harms the insert, therefore it is
prohibited. Verify that the insert is not worn.
During the operation frequently check that
the insert is intact, especially in the apical
part.
During the operation avoid prolonged
contact with the retractor or with metal
instruments in use. Do not exert excessive
pressure on the inserts during use.
5.4
Instructions for Use
After having connected all the accessories as described in
as follows:
To fill the irrigation circuit, use the PUMP
function
by
selecting
on the touch keyboard: all the other
selections will be deactivated and the
writing PUMP/FLUSH will flash;
autoclaving the handpiece, the inserts, the
torque wrench, or any other sterilisable
accessory, wait until it has completely cooled
before reuse.
inside the connectors on the handpiece
and cord must be dry. Before connecting
the handpiece to its cord, make sure that the
electrical contacts of the connector on both
sides are perfectly dry, especially after the
sterilisation cycle in an autoclave. If necessary
dry the contacts by blowing them with
compressed air.
the handpiece can rotate. When not used the
handpiece must always be put back on its
support stand.
the patient, do not perform any maintenance
tasks on the system.
1
PUMP/FLUSH
CAUTION: Contraindications. After
CAUTION: The electrical contacts
CAUTION: Owing to its conformation,
WARNING: During the intervention on
Chapter 4.3 on page 14
perio/scaler
perio/scaler
restorative
restorative
USE
proceed
23