multipiezo
8.4.1
Manual Cleaning
Manual cleaning can be carried out as an alternative to the automatic cleaning described in
Chapter 8.4.2 on page 50
8.4.1.1
Required Materials
• Enzymatic detergent at pH 6-9;
• Water;
• Container for immersion in the
enzymatic liquid;
• Ultrasonic tank;
• Clean, soft, low-lint cloths;
• Brush with soft nylon bristles;
• Syringe;
• Demineralised water
CAUTION: DO NOT USE as disinfectants::
• very alkaline products (pH > 9);
• Products containing sodium
Prepare
a
solution
detergent
with pH 6-9, according to the
a)
instructions of the manufacturer;
CAUTION: Once used, dispose of the
enzymatic detergent correctly, do not
recycle it.
a) Process validated by independent bodies with
Mectron ENZYMEC enzymatic detergent, 0.8%
v/v.
42
.
1
of
enzymatic
hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols
• Acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.