Mectron multipiezo Manual De Uso Y Mantenimiento página 30

Ocultar thumbs Ver también para multipiezo:

Publicidad

Idiomas disponibles

Idiomas disponibles

multipiezo
WARNING: Treatments that require
irrigation. Always check the operation of the
irrigation before and during use. Make sure
that the liquid exits from the insert.
Do not use the appliance if the irrigation
does not work or if the pump is faulty.
WARNING: Checking the condition of
the device before treatment. Always check
that there is no water under the appliance.
Before every treatment always check the
perfect operation of the appliance and
efficiency of the accessories. If operation
anomalies are observed do not carry out the
treatment. Contact an Authorised Mectron
Service Centre if the faults concern the
device.
WARNING: Infection control. First
use: All parts and reusable accessories (new
or returned from an Authorised MECTRON
Service Centre) are delivered in NON-STERILE
conditions and must be treated before each
use, according to the instructions provided in
Chapter 8 on page 31
. Subsequent uses:
After each treatment, clean and sterilise all
parts and reusable accessories according to
the instructions in
Chapter 8 on page 31
WARNING: Infection control. Do not
leave liquids in the tank for long periods of
time. The tank must be filled after having
cleaned and sterilised all parts, and before a
treatment. If the tank has been filled without
having used the device, it must be emptied
at the end of the day, and all parts and
accessories must be cleaned and sterilised.
WARNING: To ensure that the handpiece
cools down activate it always with irrigation
circuit installed and properly filled. To fill
the irrigation circuit always use the FLUSH
function.
CAUTION: For correct use of the
appliance, the pedal must be pressed and
the appliance started with the insert not in
contact with the part to be treated, so that
the electronic circuit is able to recognise
the best point of resonance of the insert
without interference, thus allowing optimum
performance.
22
WARNING: Before each treatment
ensure that the insert that is suitable for the
treatment is inserted on the handpiece.
Use only the Mectron torque wrench to
secure the insert to the handpiece.
WARNING: The patient must not come
into contact with the machine core unit or
with the pedal.
WARNING: Do not change the insert
whilst the handpiece is in operation in order
to avoid causing injury to the operator.
CAUTION: Pump/flush function. After
use with aggressive and non-aggressive
solutions, the tubes and handpiece must
be cleaned using the FLUSH function (See
Chapter 6 on page 27
cleaned, the crystallisation of the salts may
seriously damage the device.
CAUTION: Pump/flush function. The
"FLUSH" function must be used after each
treatment, before starting the cleaning and
sterilisation procedures.
WARNING: Do not use the device in case
of breaks or cracks on the casing.
.
WARNING: Insert Breakage and Wear.
High frequency oscillations and wear can, in
rare cases, lead to the breakage of the insert.
Deformed or damaged inserts are susceptible
to breakage during use. Broken or worn
inserts must not ever be used.
In the event of breakage check that no
fragments remain in the treated part and
at the same time use suction effectively to
remove them.
It is necessary to instruct the patient to
breathe through the nose during the
treatment, or use a dental dam, so as to avoid
to ingest fragments of broken inserts.
Check the state of wear of the insert and
its integrity before and during each use.
If a drop in performance occurs, see to its
replacement.
The state of wear of the most common
inserts (S1, S1-S, S2, S5, P2, P4, P10) can be
checked using the supplied INSERT-CARD. To
use the INSERT-CARD correctly:
• Position the insert on the INSERT-CARD
). If the tubes are not