the most direct possible path of access
to minimise any bending of the insert.
Make delicate movements. Examine
the file often for signs of wear. If the file
breaks inside the canal, avoid contact
6
PUMP/FLUSH FUNCTION
The pump/flush function allows the irrigation circuit rinse cycle to be performed.
CAUTION: Pump/flush function.
The "pump/flush" function must be used after
each treatment, before starting the cleaning
and sterilisation procedures.
Fill the tank with water, preferably
demineralised water. Connect the tank to
the appliance;
Place the handpiece, with or without
insert, above a container for the liquid
that will be released during the cleaning
cycle;
To enter cleaning mode, select "pump/
flush" on the touch keyboard: all the
other selections will be deactivated;
NOTE: The "pump/flush" mode can be
deactivated at any time by pressing the
"pump/flush" key again. The keyboard
will be reactivated and configured in the
last setting used.
PUMP/FLUSH FUNCTION
between the instrument and the broken
file to stop it from being pushed further
it. Do not apply pressure on the insert in
an axial direction.
CAUTION: If the tubes are not rinsed,
the crystallisation of the salts may seriously
damage the appliance.