multipiezo
Problem
No liquid flows out from the
insert during operation
Poor performance
Table 7 – Quick Troubleshooting
70
Possible Cause
The insert is the type without
a liquid passage
The insert is clogged
The handpiece is clogged
The irrigation level on the
screen is set to "0"
The liquid tank is empty
The tank is not properly
installed
The silicone tubes of the
pump are not correctly
installed
The peristaltic pump is worn
The insert is not correctly
tightened on the handpiece
Insert broken, worn or
deformed
Solution
Use an insert type with a
liquid passage
Unscrew the insert from the
handpiece and release the
water passage of the insert
by blowing compressed air
through it.
If the problem persists,
replace the insert with a new
one
Contact an Authorised
Mectron Service Centre
Adjust the irrigation level
Fill the tank
Properly connect the tank to
the machine core unit
Check the connections of the
tubes
Replace the peristaltic pump
(See
Chapter 9.1 on page
56
)
Unscrew the insert and screw
it back in correctly using the
Mectron torque wrench (See
Chapter 5.4 on page 23
Replace the insert with a
new one
)