multipiezo
Symbol on keyboard
Table 6 – Error Messages
NOTE: For diagnostic warnings not included
in this list, contact the technical service desk.
68
Possible cause
Peristaltic pump malfunction
The appliance has been
switched off and on again
without waiting 5 seconds
Faults on the electrical
network or excessive
electrostatic discharges or
internal faults
Incorrect start up procedure:
the appliance was started
with the pedal pressed
Solution
Check that there are no
impediments for the rotation
of the pump.
Check that the pump and
the two pipes are correctly
installed.
Turn off and wait for 5
seconds before switching on
the appliance again.
Turn off and wait 5 seconds
before switching the
appliance on again
If the warning persists,
contact an Authorised
Mectron Service Centre.
Check that the pedal is not
pressed down. If the problem
persists, disconnect the
pedal and potentially contact
an Authorised Mectron
Service Centre.