Remove the torque wrench from the
ultrasonic tank and rinse it under running
water so as to eliminate all detergent
residue.
Brush the internal and external surfaces
of the wrench with a clean brush with soft
nylon bristles, under running water.
8.4.2
Automatic Cleaning
Automatic cleaning can be carried out as an alternative to the manual cleaning described in
Chapter 8.4.1 on page 42
8.4.2.1
Required Materials
• Alkaline detergent: neodisher® FA (0.2
% v/v);
• Neutralising liquid: neodisher® Z (0.1
% v/v);
• Water;
• Metal basket;
• Adapters;
• Thermal-disinfector.
NOTE: Make sure that the accessories are
appropriately blocked in the basket and
cannot move during washing. Any blows
could damage them. Position the instruments
in a way that the water can flow through all
the surfaces, even internal.
Position the accessories in a metal basket.
Connect the relevant adapter (supplied
as an optional) to the scaler handpiece
connector and then to the water jet
cleaning connections of the thermal-
disinfector.
Repeat the same operation for the inserts
by connecting them to the appropriate
adapters supplied as optionals.
50
26
.
1
WARNING: Avoid overloading the
thermal-disinfector as this could compromise
the effectiveness of cleaning.
WARNING: On completion of the
cleaning cycle in the thermal-disinfector,
the scaler handpiece remains at the wash
temperature for a long time. Use appropriate
precautions when extracting the scaler
handpiece from the thermal-disinfector to
prevent injury to the operator.
CAUTION: Owing to its conformation, the
scaler handpiece can rotate. When not in use,
the scaler handpiece must always be placed
on its support.