Aesculap
ELAN 4 electro micro sagittal saw GA836
Perform additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
■
Material damage such as corrosion, cracks, fracturing, premature
aging, or swelling.
►
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surface and could cause corrosion
►
For further detailed information on hygienically safe and material-pre-
serving/value-preserving reprocessing, see www.a-k-i.org, link to Pub-
lications, Red Brochure – Proper maintenance of instruments.
7.3
Preparations at the place of use
►
Separate the products from each other immediately after use.
►
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
►
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
7.4
Preparation before cleaning
Applied part
►
Prior to first mechanical cleaning/disinfection process: Assemble
ECCOS holder GB086R in a suitable tray (e.g. JF222R).
►
Place the product in the correct position in the ECCOS holder GB086R,
see Fig. F.
Tools
►
Thoroughly rinse (through) the product with running, cold water.
►
Place the product on a wire basket suitable for cleaning.
8
®
7.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
►
Do not use cleaning agents that contain ace-
tone.
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
►
Do not exceed the maximum temperature of
90 °C during thermal disinfection with DI water.
►
Dry the product for at least 10 minutes at a
maximum of 120 °C.
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
Validated cleaning and disinfection procedure
Product
Validated procedure
Applied part
Manual cleaning with immer-
sion disinfection
Manual pre-cleaning with
brush and subsequent mechan-
ical alkaline cleaning and ther-
mal disinfecting
Tools
Manual pre-cleaning with
ultrasound and brush, and sub-
sequent mechanical alkaline
cleaning and thermal disinfect-
ing
Reference
see Manual clean-
ing with immer-
sion disinfection -
applied part
see Mechanical
cleaning/disin-
fecting with man-
ual pre-cleaning -
applied part
see Mechanical
cleaning/disin-
fecting with man-
ual pre-cleaning -
tools