Aesculap
ELAN 4 electro micro sagittal saw GA836
To achieve a sterility assurance level of 10
following parameters:
Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*
Sterilization
Temp.
method
Prevacuum
270 °F/275 °F
*Aesculap has validated the above sterilization cycle and has the data on
file. The validation was accomplished in an Aesculap sterile container
cleared by FDA for the sterilization and storage of these products. Other
sterilization cycles may also be suitable, however individuals or hospitals
not using the recommended method are advised to validate any alterna-
tive method using appropriate laboratory techniques. Use an FDA cleared
accessory to maintain sterility after processing, such as a wrap, pouch, etc.
7.13 Storage
Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
e.g. 2016-07
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
Note
For further information, see instructions for use of ELAN 4 electro control
unit GA800 (TA014401).
10. Technical Service
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
DANGER
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
Do not modify the product.
14
®
-6
, Aesculap recommends the
Time
Minimum drying time
4 min
20 min
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For
service
and
repairs,
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
11. Accessories/Spare parts
Art. no.
Designation
GA258SU
ELAN 4 single-use spray nozzle for sawing
GA395SU
ELAN 4 electro single-use tube set
GB086R
ELAN 4 electro ECCOS holder for micro sagittal
saw GA836
GB600
STERILIT Power Systems oil spray
GB600860
ELAN 4 electro oil spray adapter
GB692R
ELAN 4 electro rinsing device
GB698R
ELAN 4 electro rinsing adapter
GP491R–GP497R
ELAN 4 saw blade S sagittal
TA014428
Instructions for use of ELAN 4 electro micro sagit-
tal saw GA836 (A4 for ring binder)
TA014429
Instructions for use of ELAN 4 electro micro sagit-
tal saw GA836 (leaflet)