Descargar Imprimir esta página

Braun Aesculap GA836 Instrucciones De Manejo/Descripción Técnica página 12

Microsierra sagital elan 4 electro

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
Aesculap
ELAN 4 electro micro sagittal saw GA836
7.7
Mechanical cleaning/disinfecting with manual pre-cleaning - applied part
Note
The cleaning and disinfecting machine must be of tested and approved
effectiveness (e.g. FDA approval or CE mark according to DIN EN
ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
Manual pre-cleaning with a brush
Phase
Step
I
Brushes
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Follow the instructions for use of the ELAN 4 rinsing devices and rins-
ing adapters TA014447 and TA014448.
Do not clean the applied part in an ultrasound bath.
Phase I
Clean the product with a suitable cleaning brush until all discernible
residues have been removed.
Mobilize non-rigid components, such as release button, unlocking
sleeve, etc. during cleaning.
Phase II
Connect the connector for motor cable 2 with ELAN 4 electro flushing
device GB692R.
Rinse the applied part:
– For 5 min with a tap/hose or
– 3 times for 5 seconds with a water gun
10
®
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) –
RT (cold) 5
-
Water
Chemicals
quality
D–W
-
D–W
-

Publicidad

loading