INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
6.1
The Interior Light
ACTION
Press the switch to turn the interior light ON, indicated by the yellow light.
Press again and the interior light turns OFF.
6.1.1 Operating the Laminar Air Flow
ACTION
Activating the laminar fl ow.
Press one time and the fan will run at full speed, indicated by the green light (normal
model).
A fi ve digit number will run though the display showing the total hour count for the
HEPA fi lter.
Press once more and the fan will operate at reduced speed, indicated by the green light
fl ashing (standby mode).
Press one more time and the fan will switch OFF, and the time will be shown.
There will be an alarm when the fan has run for 17,000 hours, indicating that the
HEPA fi lter must be checked. When the alarm goes on the fan can still be used.
See page 47 Annual maintenance.
CAUTION
6.2
Heating System
6.2.1 Table Heating Unit
ACTION
Press one time and the HEAT will turn ON. The display shows the actual table
temperature.
Press again and the HEAT will be turned OFF.
6.2.2 Setting the Temperature
ACTION
Press the key with the symbol SP, the display will fl ash between the
SET-POINT temperature and the unit that the temperature is shown in.
Hold down the SP key and press the UP or DOWN key to change the
SET-POINT temperature. When the SET-POINT is set, release both
keys.