CAUTION
2

GENERAL OPERATIONAL INSTRUCTIONS

2.1

Pre-Operational Notes

The workstation fan must be run at normal speed for at least 15 minutes prior to working inside
the workstation.
 A green control light indicates proper operation.
 
The work chamber is to be carefully cleaned and/or disinfected (See Section 14: Maintenance).
For reliable operation it is important that the air-fl ow conditions are as undisturbed as possible.
Therefore, never overload the work chamber - insert only those utensils necessary for the actual
work.
Objects and utensils must be carefully cleaned and/or disinfected before bringing them into the
work chamber.
Necessary utensils for use during work must be placed within reach to avoid unnecessary
movement inside the workstation.
Put on necessary personal clothing for reducing particle emission from operator
(i.e. gloves, masks and general clean room clothing). Special attention should be
given to hands and lower parts of the arms, as these are the parts of the operator most likely to
emit particles near the product.
All work in the workstation must be performed with tranquil movements. Rapid arm movements in
the chamber may cause slipstreams, which will draw contaminated air into the work chamber.
The number of transports into the work chamber must be minimized.
Transport of possibly contaminated material may create airfl ows that can contaminate the
product.
2.2

Moving the Unit

The L100 is designed as a stationary unit and it should not be moved once it has been installed.
If the workstation must be moved,contact your authorised service representative.
2.3

Instruction of the Operating Personnel

These operating instructions describe the L100 Workstation series covering the following models:
L124 IVF Workstation,
L126 IVF Workstation,
L126 DUAL IVF Workstation,
L126 MP Workstation
L124 ICSI Workstation.
The L100 IVF Workstations have been manufactured in keeping with the latest technologies and
developments. It has been tested during assembly and before delivery to ensure correct operation.
However, it may present potential hazards to user, if this unit is used in conditions that lies outside of its
intentional way of use.
INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
If the equipment is used in a manner not specifi ed by this manual, the safety
of the user may be at risk and the equipment may be damaged. Always use the
equipment as outlined in this Instruction Manual.