6.11.2 Calibrate Value Zone-1 (tn-1)

The tn-1 option is for the calibration of temperature on Zone-1 (See Section 7: Diff erent Heated
Surfaces).
If there is an off set between the value shown on display and any measurements made with a high
precision external temperature sensor, this can be corrected. The new temperature reading will be
kept as the displayed value, with temperature control conducted on this basis. When the power is
disconnected default values will be restored (Calibration should only be carried by authorised
service technicians.).
6.12

Info Menu (inFo)

Follow these steps to navigate in the INFO (inFo) menu.
ACTION
Press and hold both the UP and DOWN keys for 3 seconds to
enter the menu.
First main menu uSEr will appear in the display.
Press the DOWN key until the menu inFo appears in the
display.
Press the SP key once to enter the inFo menu.
Press the UP key to get back to a previous main menu.
Press and hold both the UP and DOWN keys for 3 seconds to
exit the menu.