INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)

12.2.2 Replacing the Interior Light

1.
Disconnect the workstation power cable at
the mains.
2.
Remove the Plexiglass lamp cover, gently
grasp the light bulb by the ends, rotate the
light bulb 90° until it loosens.
3.
Carefully remove the light tube.
Replacement is the reverse of removal.
Light tube for L124:
Ø=16 mm; L= 1449 mm; 28 Watt
Light tube for L126:
Ø=16 mm; L=1449 mm; 35 Watt
It is recommended that for the L126 Workstation
that 2 persons rotate the light tube one at each end
for removal.