INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
Keyboard
SYMPTOM
Missing segment in display -
Absent or erratic function of
operation keys.
Key not working on keyboard
Microscope Light
SYMPTOM
No Microscope Light
AntiVibration System (L126MP)
SYMPTOM
Vibration
CAUSE
Failure in the PCB
Failure in keyboard
CAUSE
Light is switched off
Bad light bulb.
Faulty electrical connections
Keyboard
Blown fuse
Defective light source PCB
CAUSE
The fl oating table plate is
touching the table plate of the
workstation.
Loose tightening bolts
Grease on the springs
44
ACTION
Contact your Service
Representative for details.
Contact your Service
Representative for details.
ACTION
Press and hold the UP key
Replace the bulb-refer to
appropriate light source manual.
Check all connections to the
light source.
Contact your Service
Representative for details.
Contact your Service
Representative for details.
Contact your Service
Representative for details.
ACTION
Center the fl oating table plate so
it does not touch the rest of the
workstation.
Tighten the four securing bolts.
Clean the springs with 70%
alcohol.