7
DIFFERENT HEATED SURFACES
The L100 Series Workstations have diff erent heated surfaces.
Find the type of heated surfaces your workstation has, and fi nd out which calibration value belongs to
which heat area.
Single T845/MP (3xregulation) right side
Type: T845/MP
Size: 800 x 430 mm
Size-MP 700 x 430
Regulation: 3
The safe area is located 50 mm from the
edge. In this case there are three diff erent
temperature sensors that may need to be
calibrated. Each zone must be calibrated
separately, to calibrate "Zone 1", measure
only temperature in "Zone 1".
The red zone on the picture is "cross-heat"
zone; do not measure the temperature in
the cross-heat zone.
Dual 2xT845 (6xregulation)
Type: 2xT845
Size: 2 x 800 x 430 mm
Regulation: 6
The safe area is located 50 mm from the edge.
In this case there are three diff erent temperature sensors on each side that may need to be calibrated.
Each zone must be calibrated separately, to calibrate "Zone 1", measure only temperature in "Zone 1".The
red zone on the picture is "cross-heat" zone; do not measure the temperature in the cross-heat zone.
Zone 6
50 mm
200 mm
INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
Zone 4
Zone 5
40 mm
40 mm
200 mm
400 mm
Zone 1
50 mm
400 mm
Zone 1
50 mm
50 mm
400 mm
35
Zone 3
Zone 2
40 mm
40 mm
200 mm
200 mm
Zone 3
Zone 2
40 mm
40 mm
200 mm
200 mm
50 mm
50 mm