INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
6.2.3 Alarm
The workstations have an alarm that warns if the temperature is too high or too low. The system also has
an error mode alarm. The alarm is a visual red light on the keyboard display with an audible pulsating
alarm tone. The alarm will sound the fi rst time the unit is powered up or after a power shortage. The
alarm can be turned off by pushing the SP key.
When the heating system is fi rst activated, it will warm until the set-temperature has been reached.
During this period the alarm will not be activated.
The following messages can be shown on the display in case of alarm:
DISPLAY
NOTE: Warming up from 20°C to 37°C takes approximately 45 minutes without warming blocks on the
table. After placing an aluminum block on the table, it may take up to 30 minutes for the block
temperature to stabilize, depending on the initial block temperature.
MESSAGE
Temperature alarm on Zone-1 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-2 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-3 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-L (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-r (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-4 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-5 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on Zone-6 (See Section 7: Diff erent Heated
Surfaces).
Temperature alarm on the heat area (See Section 7: Diff erent
Heated Surfaces).
Error mode: Reset the workstation by disconnecting the power
(sound can't be muted).
16