Reset (Rest) - CooperSurgical K-SYSTEMS L100 Serie Manual De Instrucciones

Estaciones de trabajo de fiv
Tabla de contenido

Publicidad

INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)

6.12.2 Reset (rESt)

The REST function will restore all factory set values.
NOTE: The calibration value in the "StuP" menu will remain and not be reset.
NOTE: All the setup values will remain and not be reset.
Follow these steps to navigate in the Reset (rESt) option.
ACTION
Press and hold both the UP and DOWN keys for 3 seconds to
enter the menu.
Press the DOWN key until the option "inFo" appears in the
display.
Press the SP key once to enter the inFo menu.
Press the DOWN key and the option "rESt" appears in the
display.
Press and hold the SP key.
Press either the UP key or the DOWN key to reset.
When the display shows "----", release both keys. All factory
values have now been restored.
Press and hold both the UP and DOWN keys for 3 seconds to
exit the menu.
CAUTION
All the user data will be lost when resetting.
34
KEY
DISPLAY

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

K-systems l124K-systems l126K-systems dual l126K-systems l126 mpK-systems l124 icsi

Tabla de contenido