INSTRUCTION MANUAL FOR K-SYSTEMS L100 SERIES WORKSTATIONS (English)
IMPORTANT: CONNECTING THE UNIT TO A WRONG VOLTAGE WILL CAUSE SERIOUS
DAMAGES TO THE SYSTEM.

8.1.3 Gas Connections

If the workstation is connected to gas bottles, be sure to use a high quality regulator that can be adjusted
around the required 0.5-0.7 bar (7.25-10.15 psi).
If the workstation is connected to plumbed gas circuits, verify that the pressure is in the correct range.
Connect the gas to the inlet on the workstation with a suitable silicone tube (1 tube is supplied with the
unit). Contact your Customer Service Representative if this piece is missing.
The inlet should have a gas source of 5-6% premixed CO
in air.
2
37