5.3
Safe operation
Risk of infection or injury due
to aerosol formation or parti-
cles!
WARNING
Risk of injury and/or malfunc-
tion!
WARNING
Risk of injury due to applica-
tion of the product outside the
field of view!
WARNING
Note
The motor found within this Acculan 3Ti hand piece is
powered by a magnetic system. To ensure that the
motor of this powered instrument does not inadver-
tently activate, do not allow the hand piece to come in
contact with any magnetic sources (i.e. magnetic
instrument pads). Please note that having the safety
switch in the locked position does not assure that the
motor will not run inadvertently.
Adjust the cutting depth of the Acculan 3Ti
Dermatome
Note
The skin condition (e.g. due to age) of the patient must
be considered when adjusting the cutting depth (graft
thickness)!
Adjust the cutting thickness with the setting
lever 14.
Lock 15 of setting lever 14 prevents inadvertent
changes of the cutting depth setting.
Read the cutting depth (graft thickness) on dial
wheel 16 (1/10 mm gradation).
Follow the troubleshooting advisories/procedure,
see Troubleshooting list.
Use suitable protection
(such as waterproof pro-
tective clothing, face
mask, safety gases, suc-
tion).
Always carry out a func-
tion check prior to using
the product.
Only use the product under
vision.
Adjust the cutting width of the Acculan 3Ti
Dermatome
Risk of injury/cuts by the der-
matome blade if the flaps are
opened incorrectly!
WARNING
Note
The width of the skin graft will be approx. 2 mm less
than the cutting width set by the flaps!
Set the required cutting width by opening the
appropriate number of flaps 12.
Safely lift flaps 12 by pressing flap tabs 11, see
Fig. 12.
12
11
Fig. 12 Safely lifting the flaps
Completely open flaps 12 until they engage in their
end positions (flap width = 8 mm).
Always use the flap tabs
for lifting the flaps.
en