Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfecting machine without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate
rinse
IV
Thermal disin-
fection
V
Drying
DW:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alcaline
►
Insert the product in its proper position in the
ECCOS fixation. Insert the battery compartment
over the clamp and swivel device into the ECCOS
fixation so that the trigger 4 points upwards.
►
Ensure that the blade cover 6 is kept open through
the entire processing procedure.
►
Dismantle the flap rod 18 for cleaning. Store flap
rod 18, flaps 12 and nut 2 separately in tray.
►
After automatic cleaning/disinfection, check visible
surfaces for residues and repeat the cleaning/disin-
fection process as needed.
D
t
Water
[°C/°F]
[min]
quality
<25/77
3
DW
55/131
10
FD–W
> 10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
■
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
working solution 0.5%
– pH ~ 11*
-
-
mind. 10 min at max. 120 °C
6.10 Inspection, maintenance and
checks
Note
Aesculap recommends lubricating the flap rod, flaps,
sliding surfaces and other moving parts, (e.g. trigger,
cover lids) with STERILIT-Power-Systems oil spray
GB600 or the STERILIT-Power Systems drip lubricator
GA059 prior to every sterilization. Remove excess oil
with a lint-free wipe.
►
Allow the product to cool down to room tempera-
ture.
►
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is: clean, functional, and
undamaged.
►
Check the product for any damage, abnormal run-
ning noise, overheating or excessive vibration.
►
Set aside the product if it is damaged.
en
19