Operate Acculan 3Ti Dermatome
Note
The low howling noise heard during startup of the
Acculan 3Ti Dermatome is normal for this product!
Press trigger 4.
The saw runs at the continuously controlled oscil-
lation frequency.
Taking skin grafts
To familiarize with the operation of the Acculan 3Ti
Dermatome prior to its first surgical application, the
user should carry out some test cuts on a specimen.
Stretch flabby skin.
With the cutting surface in contact with the skin to
be cut, smoothly push forward the dermatome 1,
applying gentle pressure. Be careful not to cant
dermatome 1.
For very thin grafts, use a forceps, from time to
time, to lay up the resected portion of the graft on
surface 13.
Turn off the motor and pull the freshly cut skin
graft out of the device
- or -
Lower the handle and cut off the skin graft with
dermatome 1.
14
6.
Validated reprocessing pro-
cedure
6.1
General safety information
Note
Adhere to national statutory regulations, national and
international standards and directives, and local, clini-
cal hygiene instructions for reprocessing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), sus-
pected CJD, or possible variants of CJD, observe the rel-
evant national regulations concerning the reprocessing
of products.
Note
Mechanical reprocessing should be favored over man-
ual cleaning as it gives better and more reliable results.
Note
It should be noted that successful reprocessing of this
medical device can only be guaranteed following prior
validation of the reprocessing method. The opera-
tor/reprocessor is responsible for this.
The recommended chemical was used for validation.
Note
If there is no final sterilization, then a virucidal disinfec-
tant must be used.
Note
For the latest information on reprocessing and material
compatibility see also the Aesculap extranet at
https://extranet.bbraun.com
The validated steam sterilization procedure was carried
out in the Aesculap sterile container system.