en
6.6
Single-use products
Dermatome blade GB228R
Risk of infection for patients
and/or users and impairment
of product functionality due
WARNING
to reuse. Risk of injury, illness
or death due to contamination
and/or impaired functionality
of the product!
►
6.7
Cleaning/disinfection
Product-specific safety instructions for the
reprocessing procedure
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
CAUTION
sive temperatures!
►
►
►
►
►
►
16
Do not reprocess the der-
matome blade GB228R.
Use cleaning and disinfect-
ing agents according to the
manufacturer's instruc-
tions which
be approved for plastic
material and high-
grade steel,
do not attack softeners
(e.g. in silicone).
Do not use cleaning agents
that contain acetone.
Observe specifications
regarding concentration,
temperature and exposure
time.
Do not exceed the maxi-
mum temperature of 60 °C
during chemical cleaning
and/or disinfection.
Do not exceed the maxi-
mum temperature of 96 °C
during thermal disinfection
with FD water.
Dry the product for at least
10 minutes at a maximum
of 120 °C.
Damage or destruction of the
batteries due to processing!
►
CAUTION
Note
The indicated drying time is a guide time only. It must be
checked taking into account the specific conditions (e.g.
load) and if applicable adjusted.
Protect battery from mois-
ture.