Descargar Imprimir esta página

B.Braun Aesculap Acculan 3Ti Instrucciones De Manejo/Descripción Técnica página 14

Ocultar thumbs Ver también para Aesculap Acculan 3Ti:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 70
en
Protection against inadvertent activation
To prevent inadvertent activation of dermatome 1
while replacing the dermatome blade, trigger 4 must
be locked.
To lock trigger 4: Move safety catch 5 to its OFF
position, see Fig. 10.
Trigger 4 is locked. Dermatome 1 cannot be acti-
vated.
To unlock trigger 4: Move safety catch 5 to its ON
position, see Fig. 10.
Trigger 4 is unlocked. Dermatome 1 can be acti-
vated.
4
5
Fig. 10 Engaging the safety catch
Intraoperative storage
Note
The Eccos holder can be used for intraoperative storage.
The dermatome can not be processed in this position.
For the correct position, see Automatic cleaning/disin-
fection with manual pre-cleaning!
12
1
Engage safety catch 5 against inadvertent activa-
tion, see Protection against inadvertent activation.
Fold back the clamp of the ECCOS holder.
Place dermatome 1 in the ECCOS holder, see
Fig. 11.
Fig. 11 Store the dermatome ready for operation in
the ECCOS holder
5.2
Function checks
Note
The function checks must be carried out prior to each
surgical application of the unit and after each intraop-
erative battery change!
Follow the advisories/procedure for inspection,
maintenance and checks, see Inspection, mainte-
nance and checks.
Ensure that a dermatome blade is correctly seated
in the dermatome.
Make certain that a battery has been installed.
Check that the roll pin is securely seated.
Check that the dermatome blade is securely seated.
Briefly activate the dermatome at maximum oscil-
lation frequency.

Publicidad

loading

Este manual también es adecuado para:

Ga670