INSTRUCTIONS FOR USE
11. Grasp the boot release lever and pull up to unlock the stirrup.
12. Move the stirrup along the main support bar until the top end of calf section of the
stirrup is approximately at the patient's mid-calf.
13. Release the boot release lever to lock the stirrup in place.
14. Do Step 2 through Step 13 on the other side of the table.
3.2.3

Pre-Use Function Checks

Make sure that the Yellofins Apex™ is in good working order. Do the checks that follow:
• Grasp and firmly squeeze the lithotomy handle. Make sure that
the Yellofins Apex™ moves freely through the full range of
lithotomy adjustment, and ab/ad-duction. Release the
lithotomy handle.
• Grasp the boot release lever and pull up to unlock the stirrup.
Make sure that the stirrup does not release before the boot
release lever is fully engaged. Make sure that the stirrup moves
freely through the full range of adjustment. Make sure that the
stirrup moves freely along the main support tube. Release the
boot release lever.
• Apply moderate pressure to the lithotomy handle in all directions. Make sure that the
Yellofins Apex™ does not move.
• Apply moderate pressure to the stirrup boot in all directions. Make sure that the stirrup
boot does not move.
• Make sure that the straps engage securely and do not disengage when light force is
applied.
Document Number: 80028475
Version: A
Page 17
English
Issue Date: 26 November 2020
Ref Blank Template: 80025117 Ver. F