combi touch
8.4.1.3
Inserts
Place the insert in a clean container,
in a horizontal position. Add enough
enzymatic
detergent
completely cover it.
Leave
the
insert
enzymatic
detergent
10 minutes at ambient temperature
(22 °C ±2 °C);
While soaking in the enzymatic solution,
gently brush all the surfaces until all
visible soil has been removed.
Use a clean brush with soft nylon bristles
for the external surfaces, and a clean pipe
cleaner with soft nylon bristles for the
internal cavities and grooves.
NOTE: Thoroughly brush all the
following parts of the insert for about 20
seconds:
• Through-holes and internal
channels;
• Hard-to-clean areas such as
sharp edges and in particular, the
interstices between the cutting
cusps;
• Internal cavities, grooves and
fissures.
Remove the insert from the enzymatic
detergent solution.
Thoroughly rinse and brush all the
surfaces of the insert (see previous point)
under running water for at least 10
minutes;
82
solution
to
to
soak
in
the
solution
for
12
13
14