combi touch
Repeat points 20 to 23 another 4 times.
The number of cycles run must guarantee
the complete drainage of the external
irrigation circuit. In any case, check that
liquids are no longer coming out of the
quick coupling on the cord of the air-
polishing handpiece;
Disconnect the syringe and repeat points
13 to 24 another 2 times;
Disconnect the syringe and fill it with
distilled water;
NOTE: Use of distilled water is
recommended for long periods of
non-use. If the device is used after the
disinfection procedure, use common
running water.
Reconnect the syringe to the end
of the hose from which it had been
disconnected.
Press the "flush" button on the touch
keyboard (see
Chapter 6.2 on page 43)
58
24
25
26
27
28
.
refill