11.3
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment
Even if compliant with standard IEC/EN
60601-1-2, combi touch may interfere with
other devices in the vicinity. combi touch
must not be used near or stacked with other
appliances. However, if this is necessary, the
correct operation of the equipment in that
configuration, and of all equipment, must be
checked and monitored before starting the
intervention.
WARNING: Portable and mobile radio
communication equipment may influence the
correct operation of the appliance.
11.3.1 Guide and Manufacturer's Declaration - Electromagnetic
Emissions
combi touch is designed to operate in the electromagnetic environment specified below. The
purchaser or user of combi touch should ensure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
WARNING: Contraindications.
Interference from other equipment
An electrosurgical scalpel or other
electrosurgical units positioned near the
combi touch appliance may interfere with the
correct operation of the appliance itself.
WARNING: The appliance requires
particular EMC precautions and must be
installed and put into service according to the
EMC information provided in this chapter.
WARNING: The use of cables and
accessories not supplied by MECTRON may
adversely affect the EMC performances.
Compliance
Electromagnetic Environment Guidance
combi touch uses RF energy only for its
internal operation. Therefore, its RF emissions
Group 1
are very low and probably do not cause any
interference with nearby electronic devices.
Class B
combi touch is suitable for use in all
buildings, including domestic buildings, and
Class A
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Compliant