Mectron combi touch Manual De Uso Y Mantenimiento página 227

Ocultar thumbs Ver también para combi touch:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Leave to soak for 10 minutes at 22°C
±2°C. This procedure reduces the amount
of blood, protein and mucous present on
the air-polishing handpiece;
After 10 minutes of immersion in the
enzymatic solution, gently brush the
surface of the air-polishing handpiece
using the brush with soft nylon bristles
for at least 20 seconds, and eliminate
all visible residue. Thoroughly clean the
hard-to-reach areas such as the edges,
recesses, and joints;
Rinse the air-polishing handpiece under
hot running water (40 °C ±5 °C) for about
10 minutes, moving it slightly to allow the
water to reach the entire surface;
Place the air-polishing handpiece in
the ultrasonic tank immersed in the
enzymatic detergent solution for at least
10 minutes at 22°C ±2°C;
CLEANING AND STERILISATION
29
30
31
32
87