combi touch
Press and release the pedal. When the
"flush" cycle begins, start injecting the
remaining disinfectant with the syringe.
During the "flush" cycle, inject the
10ml of remaining disinfectant.
The "flush" cycle on the "Ultrasound" part
lasts about 23 seconds;
NOTE: The "flush" cycle on the
"Ultrasound" part can be interrupted at
any time by pressing and releasing the
pedal. Before running the "flush" cycle
on the "Ultrasound" part check that
the left-hand knob in the front part of
the device (Ref. 5 inside cover) is open.
At the end of the "flush" cycle, stop
injecting the liquid with the syringe.
CAUTION: Do not press too hard
on the syringe. The irrigation circuit
hoses are quite small and it is normal for
injection to occur slowly.
NOTE: If the 10ml of disinfectant
are injected before the flush cycle is
completed, wait for it to finish. If the
cycle ends before the full 10ml have
been injected, repeat the flush cycle until
the 50ml have been completely injected.
Disconnect the syringe and aspirate air;
64
13
14