Re-insert the previously cleaned water
filter (see point 4 of the procedure) in its
seat - Ref. 1 -, screw in the original and
previously disinfected male coupling (see
points 5 and 6 of the procedure) - Ref. 2
- and reconnect the quick coupling of the
external water circuit - Ref. 3.
8.1.1.3
Procedure - Scaler Handpiece
WARNING: Switch the appliance
off. Switch the appliance off using the
switch (Ref. 1), disconnect the power
supply cable from the wall socket
(Ref. 2) and machine core unit (Ref. 3)
before carrying out the cleaning and
sterilisation tasks.
Carry out steps 2 to 10 described in
Chapter 8.1.1.2 on page 51;
Extract the quick coupling on the cord of
the scaler handpiece and place it above
a container or sink to catch the liquid
released during the procedure;
CAUTION: Before connecting and
using the scaler handpiece, make sure
that the electrical contacts on both
the scaler handpiece and its cord are
perfectly dry.
60
33
3
1
2
3
1
2
1
3
2