Mectron combi touch Manual De Uso Y Mantenimiento página 196

Ocultar thumbs Ver también para combi touch:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

combi touch
Press the "flush" button on the touch
keyboard (see
Chapter 6.2 on page 43
Press and release the pedal. When the
"flush" cycle begins, start injecting the
remaining disinfectant with the syringe.
During the "flush" cycle, inject the
10ml of remaining disinfectant.
The "flush" cycle on the "Polishing" part
lasts about 40 seconds;
NOTE: The "flush" cycle on the
"Polishing" part cannot be interrupted.
Before running the "flush" cycle on the
"Polishing" part check that the right-
hand knob in the front part of the device
(Ref. 9 inside cover) is open. At the end
of the "flush" cycle, stop injecting the
liquid with the syringe.
CAUTION: Do not press too hard on
the piston of the syringe. The irrigation
circuit hoses are quite small and it is
normal for injection to occur slowly.
NOTE: If the 10ml of disinfectant
are injected before the flush cycle is
completed, wait for it to finish. If the
cycle ends before the full 10ml have
been injected, repeat the flush cycle until
the 10ml have been completely injected.
56
18
);
19
refill

Publicidad

Tabla de contenido
loading

Tabla de contenido