Idiomas disponibles

hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols
• Acetone;
8.3

Cleaning the Tank and Cap

The following procedure must be carried out on the tank and cap of the device.
8.3.1

Preparation

1.
Disconnect the tank from the machine
core unit (see
Chapter 7 on page 45
2.
Unscrew the cap from the tank.
8.3.2

Required Materials

• Water;
• Cleaning solution (pH 6-9);
• Clean, soft, low-lint cloth;
• Demineralised water.
CAUTION: DO NOT USE as disinfectants::
• Very alkaline products (pH > 9);
• Products containing sodium
hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
8.3.3

Cleaning Method

Thoroughly rinse the inside and outside
of both the tank and cap under running
water;
CLEANING AND STERILISATION
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.
CAUTION: Do not sterilise the tank and
);
cap in an autoclave. They could be damaged.
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols
• Acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.