11.3.3.3 Parts Accessible To The Input / Output Signals - Mectron combi touch Manual De Uso Y Mantenimiento

Ocultar thumbs Ver también para combi touch:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

combi touch
11.3.3.3 Parts Accessible to the Input / Output Signals
combi touch is designed to operate in the electromagnetic environment specified below.
The purchaser or user of combi touch should ensure that it is used in such an environment.
Essential EMC
Phenomenon
standard or test
Electrostatic
discharges
IEC 61000-4-2
(ESD)
e)
Electrical fast
transient/burst
IEC 61000-4-4
b) f)
Pulses
Common
IEC 61000-4-5
mode
a)
Conductive
disturbances
IEC 61000-4-6
induced by RF
fields
b) d) g)
a) This test applies only to output lines connected
directly to the external cables.
b) SIP/SOPS with maximum cable length less than 3 m
are excluded.
c) The tests can be performed with other modulation
frequencies identified by the RISK MANAGEMENT
PROCESS.
d) The calibration of the current injection terminals must
be performed in a system at 150 Ω.
e) The connectors have to be tested in accordance
with paragraph 8.3.2 and Table 4 of the standard IEC
61000-4-2:2008. For insulated connector housings,
perform the air discharge test on the connector
housing and pins using the probe with the rounded
tip of the ESD generator, with the exception that only
the connector pins that are tested are those that
can be reached or touched, under the conditions
of INTENDED USE, by the standard probe shown in
Figure 6 of the general standard, applied in a bent or
straight position.
f) It is necessary to use the capacitive coupling.
g) If an ISM or amateur radio band is not present among
110
Immunity test values
method
±8 kV on contact
±2 kV, ±4 kV, ±8 kV,
±15 kV in air
±1 kV on contact
100 KHz repetition
frequency
± 2kV
3 V
h)
0.15 MHz - 80 MHz
6 V
in the ISM bands
h)
between 0.15 MHz and
80 MHz
80% AM at 1 KHz
Environment Guidance
The floor must be made of
wood, concrete or ceramic
tiles. If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
i)
distances, calculated from the
c)
equation applicable at the
frequency of the transmitter.
the frequency samples, as appropriate, an additional
test frequency has to be used in the ISM band or in
the amateur radio band. This applies to each ISM and
amateur radio band within the specified frequency
range.
h) R.M.S., applied before modulation.
i) The ISM bands (industrial, scientific and medical)
between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz
to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The
amateur radio bands between 0.15 MHz and 80 MHz
are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz
to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15
MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz,
21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0
MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
Electromagnetic

Publicidad

Tabla de contenido
loading

Tabla de contenido