combi touch
At the end of the disinfection procedure,
remove the syringe (Ref. 1), the hose (Ref.
2) and unscrew the male coupling used
(Ref. 3).
Re-insert the previously cleaned water
filter (see point 4 of the procedure) in its
seat - Ref. 1 -, screw in the original and
previously disinfected male coupling -
Ref. 2 - and reconnect the quick coupling
of the external water circuit - Ref. 3.
8.1.2
Disinfecting the Tank Irrigation Circuit
After having run the flush function (see
irrigation circuit (see
cleaning procedures, proceed to disinfect the tank irrigation circuit.
To disinfect the irrigation circuit, proceed as follows:
8.1.2.1
Required Materials
• CIDEX® OPA disinfectant solution;
• Distilled water.
68
Chapter 6 on page 40
Chapter 8.1.1 on page 51)
28
1
29
3
), disinfected the external
, and before carrying out the subsequent
3
2
1
2