or 490 Auto-reader having software version 4.0.0
or greater or 490M Auto-reader with a new positive
control.
With processed 1295 BIs, a final negative reading
(- symbol on the display) indicates an acceptable
sterilization process. A positive (+ symbol on the display)
result for a processed 1295 BI indicates a sterilization
process failure. Act immediately on any positive results
for processed BIs. Determine the cause of the positive
BI following facility policies and procedures. 3M
recommends retesting the sterilizer according to your
facility policy before processing additional loads.
Optional Visual pH Color Change Result
The 1295 BI is normally discarded after the fluorescent
result has been recorded. If, however, special studies
are desired, 1295 BIs may be further incubated for a
visual pH color change result. In the case of the positive
control BI, a yellow color change of the growth media
will appear within 24 hours. Any observation of a yellow
color within the vial indicates a positive result.
In the case of a processed 1295 BI, a media color
change from purple to yellow indicates a sterilization
process failure. A negative pH color change result, i.e.,
media remains purple, can be assessed at 7 days. To
avoid media dry-out, it is recommended that the 1295
BI be transferred to a humidified incubator operating at
60ºC after the fluorescent result has been recorded.
Indicateur biologique à lecture rapide 3M™ Attest™ 1295

Description du produit

L'indicateur biologique à lecture rapide 3M™ Attest™
1295 (muni d'un capuchon rose et ci-après désigné sous
le nom de 1295 BI) est un indicateur biologique
autonome spécialement conçu pour un contrôle de
Storage/Disposal
Storage
• Store 1295 BIs in the original resealable foil
pouch under normal room conditions: 59-86ºF
(15-30ºC).
• Do not store 1295 BIs near sterilants or other
chemicals.
Disposal
Dispose of used 1295 BIs according to your health
care facility policy. You may wish to steam sterilize
any positive biological indicators at 132°C (270°F)
for 4 minutes or at 275°F (135°C) for 3 minutes in a
dynamic-air-removal steam sterilizer prior to disposal.
For processed BIs observed to have a broken media
ampoule after the sterilization process, or if the ink of
the CI appears smeared or runny (if included) leave the
BI and CI in the sterilization pouch and place the pouch
in packaging compatible with steam sterilization and
steam sterilize per the parameters above.
For further information, please contact your local 3M
representative or contact us at 3M.com and select your
country.
3M and Attest are trademarks of 3M
STERRAD and ALLClear are registered trademarks of
Advanced Sterilization Products
V-PRO is a registered trademark of Steris Corporation.
routine rapide et fiable des processus de stérilisation par
vaporisation de peroxyde d'hydrogène lorsqu'il est
utilisé en association avec l'Auto-reader 3M™ Attest™
490H ou avec l'Auto-reader 3M™ Attest™ 490 doté de
la version logicielle 4.0.0 ou ultérieure (ci-après désigné
sous le nom d'Auto-reader 490H ou Auto-reader 490
23
b

Publicidad

Tabla de contenido
loading

Tabla de contenido