3.0 BASIC OPERATION
3.1 Control Operation
Refer to the model/serial number label on the head end of the table for input voltage
requirements. Plug the power cord into the power cord receptacle, located at the head end
of the table and into a properly grounded receptacle. Turn on the on/off power switch. The
green light in the power switch illuminates, indicating that AC power is applied to the table.
This switch functions as a combination on/off power switch, circuit breaker and pilot light.
NOTE: A startup self-diagnostic routine is initiated each time the table is plugged in and
turned on.
3.2 Caster Floor Locks
Prior to transferring a patient onto the table and at all times when the table is in use, ensure
all four (4) independent, manual caster floor locks are in the locked position. To engage the
caster lock, step on the foot pad and depress to the floor. To unlock, slide foot under pad
and lift to disengage lock.
LOCK
Figure 5: Floor lock in the unlocked position
NOTE: The casters and locks installed on your hana
those depicted above but they will operate in the same manner that is described.
3.3 Moving The Table
After all four (4) floor locks are unlocked, the table can be rolled for relocation. It is best that
the table be rolled with the foot end pointed in the direction of travel. The 6875 hana
Surgery Table is heavy and a minimum of two people should move it, one at the head end,
one at the foot end, and care should be taken to control it when rolling.
If the table is allowed to roll too fast, it may be difficult to stop or turn. Impact of the
table top with a stationary object may cause serious damage to the table top. If an
impact occurs the table must be visually inspected for visible damage, and a Function
Check must be performed (see section 5.0 of this manual). If damage is discovered or
the table does not pass the Function Check call your Mizuho OSI service
representative.
MIZUHOSI 2009
Location of external
ground stud
UNLOCK
Figure 6: Floor lock in the locked position
®
Surgery Table may look different than
WARNING:
11
®