5.4 Press and hold TRENDELENBURG button. Head end of tabletop becomes lower
than the foot end
5.5 Press and hold LEFT LATERAL ROLL button. Observe that the tabletop tilts to the
left.
5.6 Press and hold RIGHT LATERAL ROLL button. Observe that the tabletop tilts to the
right.
5.7 Move tabletop out of level in both tilt and trendelenburg by 5 degrees and raise table
at least 2 inches from lowest position. Press and hold RETURN TO LEVEL button.
The tabletop first levels roll and trendelenburg and then after a three-second delay,
the tabletop moves down to its lowest height position. Observe that the table top is
level and in its lowest height position.
6. Control panel check:
6.1 Repeat steps 5.1 through 5.7 using the control panel buttons.
7. Femur lift control and foot pedal check:
7.1 With femur lift connected with electrical connectivity, the corresponding light on the
control panel should illuminate indicating a good connection. See section 8.3 for
proper installation of femur lift and foot pedal.
7.2 With femur lifts and foot pedal connected and connectivity indicator lights green,
position the control knob located in section 1 of the control panel (see figure 4) to
either the left femur lift or right femur lift. Test femur lift foot pedal function for
operation of the indicated femur lift. Repeat for opposite side.
MIZUHOSI 2009
17
NW0508 Rev. D